American Society of Anesthesiologists Summary of Anesthesia Machine Checkout Recommendations

American Society of Anesthesiologists Summary of Anesthesia Machine Checkout Recommendations
Item to be completed Responsible party
To be completed daily
Item #1: Verify that auxiliary oxygen cylinder and self-inflating manual ventilation device are available and functioning Provider and technician
Item #2: Verify that patient suction is adequate to clear the airway Provider and technician
Item #3: Turn on anesthesia delivery system and confirm that AC power is available Provider or technician
Item #4: Verify availability of required monitors, including alarms Provider or technician
Item #5: Verify that pressure is adequate on the spare oxygen cylinder mounted on the anesthesia machine Provider and technician
Item #6: Verify that the piped gas pressures are ≥50 psig Provider and technician
Item #7: Verify that vaporizers are adequately filled and, if applicable, that the filler ports are tightly closed Provider or technician
Item #8: Verify that there are no leaks in the gas supply lines between the flowmeters and the common gas outlet Provider or technician
Item #9: Test scavenging system function Provider or technician
Item #10: Calibrate, or verify calibration of, the oxygen monitor, and check the low oxygen alarm Provider or technician
Item #11: Verify that carbon dioxide absorbent is not exhausted Provider or technician
Item #12: Breathing system pressure and leak testing Provider and technician
Item #13: Verify that gas flows properly through the breathing circuit during both inspiration and exhalation Provider and technician
Item #14: Document completion of checkout procedures Provider and technician
Item #15: Confirm ventilator settings and evaluate readiness to deliver anesthesia care (anesthesia time out) Provider
To be completed prior to each procedure
Item #2: Verify that patient suction is adequate to clear the airway Provider and technician
Item #4: Verify availability of required monitors, including alarms Provider or technician
Item #7: Verify that vaporizers are adequately filled and, if applicable, that the filler ports are tightly closed Provider
Item #11: Verify that carbon dioxide absorbent is not exhausted Provider or technician
Item #12: Breathing system pressure and leak testing Provider and technician
Item #13: Verify that gas flows properly through the breathing circuit during both inspiration and exhalation Provider and technician
Item #14: Document completion of checkout procedures Provider and technician
Item #15: Confirm ventilator settings and evaluate readiness to deliver anesthesia care (anesthesia time out) Provider
AC: alternating current; psig: pounds per square inch gauge.
Reproduced with permission from: Riutort KT, Eisenkraft JB. The Anesthesia Workstation and Delivery Systems for Inhaled Anesthetics. In: Clinical Anesthesia, 7th ed, Barash PG, Cullen BF, Stoelting RK, et al. (Eds), Lippincott Williams & Wilkins, Philadelphia 2013. Copyright © 2013 Lippincott Williams & Wilkins. www.lww.com.
Graphic 101563 Version 6.0