ﺑﺎﺯﮔﺸﺖ ﺑﻪ ﺻﻔﺤﻪ ﻗﺒﻠﯽ
خرید پکیج
تعداد آیتم قابل مشاهده باقیمانده : 3 مورد
نسخه الکترونیک
medimedia.ir

MarginProbe

MarginProbe
Study Intervention Positive margin rate Re-excision rate
Arm n Percent p value Percent p value
Schnabel et al[1] Device 298 30.9 0.008 19.8 0.097
SOC 298 41.6 25.8
Sebastian et al[2] Device 165 9.7 <0.0001
SOC 186 25.8
Thill et al[3] Device 42  –  – 17 0.018
SOC 67  – 39
Allweis et al[4] Device 143  – –  12.6 0.098
SOC 150  – 18.6
SOC: standard of care.
References:
  1. Schnabel F, Boolbol SK, Gittleman M, et al. A randomized prospective study of lumpectomy margin assessment with use of MarginProbe in patients with nonpalpable breast malignancies. Ann Surg Oncol 2014; 21:1589.
  2. Sebastian M, Akbari S, Anglin B, et al. The impact of use of an intraoperative margin assessment device on re-excision rates. SpringerPlus 2015; 4:198.
  3. Thill M, Dittmer C, Baumann K, et al. MarginProbe®--final results of the German post-market study in breast conserving surgery of ductal carcinoma in situ. Breast 2014; 23:94.
  4. Allweis TM, Kaufman Z, Lelcuk S, et al. A prospective, randomized, controlled, multicenter study of a real-time, intraoperative probe for positive margin detection in breast-conserving surgery. Am J Surg 2008; 196:483.