Operating parameters for continuous flow left ventricular assist devices

Operating parameters for continuous flow left ventricular assist devices
LVAD trade name (rotor design)

Pump speed (RPM)

Usual clinical range (full range)

Pump flow (L/minute) Pulsatility index (flow pulsatility) Pump power (W)
HeartMate III (centrifugal) 5000 to 6000  (3000 to 9000) 4 to 6 3.5 to 5.5 4.5 to 6.5
HeartWare HVAD System (centrifugal) 2800 to 3200 (1800 to 4000) 4 to 6 2 to 4 3 to 7
HeartMate II (axial) 8000 to 10,000 (6000 to 15,000) 4 to 7 5 to 8 5 to 8
LVAD: left ventricular assist device; RPM: revolutions per minute; W: watts.
Reproduced from: Dalia AA, Cronin B, Stone ME, et al. Anesthetic Management of Patients With Continuous-Flow Left Ventricular Assist Devices Undergoing Noncardiac Surgery: An Update for Anesthesiologists. J Cardiothorac Vasc Anesth 2018; 32:1001. Table used with the permission of Elsevier Inc. All rights reserved.