Crosstalk B

Crosstalk B
If the ventricular sensing circuit senses activity during the crosstalk sensing window, a ventricular pacing artifact is delivered early, usually at 100-110 milliseconds after the atrial event. This has been referred to as "ventricular safety pacing," "110 millisecond phenomenon," and "nonphysiologic AV delay."
AV: atrioventricular
From: Hayes DL, Wang PJ, Asirvatham SJ, Friedman PA. Timing cycles. In: Cardiac Pacing, Defibrillation and Resynchronization: A Clinical Approach, 3rd ed, Hayes DL, Asirvatham SJ, Friedman PA (Eds), Wiley-Blackwell, West Sussex 2013. Copyright © 2013 Mayo Foundation for Medical Education and Research. Reproduced with permission of John Wiley & Sons Inc. This image has been provided by or is owned by Wiley. Further permission is needed before it can be downloaded to PowerPoint, printed, shared or emailed. Please contact Wiley's permissions department either via email: [email protected] or use the RightsLink service by clicking on the 'Request Permission' link accompanying this article on Wiley Online Library (https://onlinelibrary.wiley.com/).