Personal CGM devices commonly used in the United States*

Personal CGM devices commonly used in the United States*
  FreeStyle Libre 2/FreeStyle Libre 2 Plus FreeStyle Libre 3/FreeStyle Libre 3 Plus Dexcom G6 Dexcom G7 Dexcom Stelo Medtronic Guardian 3 Medtronic Guardian 4 Eversense E3 Eversense E365
Insertion site Arm Arm Abdomen Abdomen or arm Arm Abdomen or arm Abdomen or arm Arm (implanted subcutaneously) Arm (implanted subcutaneously)
Duration of sensor wear 14 days/15 days 14 days/15 days 10 days 10 days 15 days 7 days 7 days 180 days 365 days
Calibration required? No No No No No Yes, every 12 hours (3 to 4 times recommended) No Yes, every 12 hours for the first 21 days of use, then once daily Yes, every 12 hours for the first 14 days of use, then once weekly
Warm-up time 1 hour 1 hour 2 hours 30 minutes 30 minutes 2 hours 2 hours 24 hours 24 hours
Method of insertion Self-insertion Self-insertion Self-insertion Self-insertion Self-insertion Self-insertion Self-insertion Requires office-based procedure Requires office-based procedure
Optional or customized alarms for glucose values Low and high Low and high Predictive low and high, rate of change, low and high Predictive low and high, rate of change, low and highΔ No (not recommended for people using insulin) Predictive low and high, rate of change, low and high Predictive low and high, rate of change, low and high Predictive low and high, rate of change, low and high Predictive low and high, rate of change, low and high
Programmed/mandatory alarms for glucose values None Urgent low Urgent low Urgent low None (not recommended for people using insulin) Urgent low Urgent low Urgent low Urgent low
Insulin pump integration? No/Tandem t:slim X2 mylife YpsoPump Tandem t:slim X2, Omnipod 5, and mylife YpsoPump Tandem t:slim X2 and Mobi; Omnipod 5 No Medtronic MiniMed 670G, 770G, 780G Medtronic MiniMed 780G No No
Share feature for health care providers LibreView LibreView Clarity Clarity Clarity CareLink CareLink Eversense Data Management System (DMS) Pro Eversense Data Management System (DMS) Pro
Intermittent scanning ("flash") or real-time? Intermittent scanning/Real-time Real-time Real-time Real-time Real-time Real-time Real-time Real-time Real-time

CGM: continuous glucose monitoring.

* Listed devices are intended for personal use. Professional CGM devices (practice-owned) include the FreeStyle Libre Pro, the Dexcom G6 Pro, and the Medtronic iPro2.

¶ Fingerstick glucose measurements are also required to determine prandial insulin dosing.

Δ The Dexcom G7 has a "delay first alert" setting for high glucose. The alarm may be delayed until glucose remains elevated for a specified duration.

◊ All listed devices also have compatible smartphone apps that allow sharing of personal CGM data with followers (eg, family, friends, caregivers).

Graphic 140874 Version 2.0