Vagus nerve stimulator: adjustable parameters and initial settings

Vagus nerve stimulator: adjustable parameters and initial settings
  Initial settings Adjusted settings Comment
On time 30 seconds    
Off time 5 minutes 5, 3, 1.8, 1.1 minutes Decreasing off time, (increasing the duty cycle) may improve efficacy, but may decrease tolerability and battery life
Output current 0.25 mA 0.25 mA to 3.5 mA Increased current may improve efficacy, but may decrease tolerability
Frequency 30 Hz 20 to 30 Hz Decreasing frequency may improve tolerability without affecting efficacy
Pulse width 500 microseconds 250, 500 microseconds Decreasing pulse width may improve tolerability without affecting efficacy
Magnet current 0.5 mA    
Magnet duration 60 seconds    
Magnet pulse width 500 microseconds    
Graphic 59936 Version 2.0