Initial settings | Adjusted settings | Comment | |
On time | 30 seconds | ||
Off time | 5 minutes | 5, 3, 1.8, 1.1 minutes | Decreasing off time, (increasing the duty cycle) may improve efficacy, but may decrease tolerability and battery life |
Output current | 0.25 mA | 0.25 mA to 3.5 mA | Increased current may improve efficacy, but may decrease tolerability |
Frequency | 30 Hz | 20 to 30 Hz | Decreasing frequency may improve tolerability without affecting efficacy |
Pulse width | 500 microseconds | 250, 500 microseconds | Decreasing pulse width may improve tolerability without affecting efficacy |
Magnet current | 0.5 mA | ||
Magnet duration | 60 seconds | ||
Magnet pulse width | 500 microseconds |