Minimum frequency of cardiac implantable electronic device (CIED) follow-up

Minimum frequency of cardiac implantable electronic device (CIED) follow-up
Type Frequency Delivery method
Immediate post-implant check Within 72 hours In person
Early post-implant check 2 to 12 weeks In person
Routine PPM/CRT-PPM check Every 3 to 12 months In person or remote
Routine ICD/CRT-D Every 3 to 6 months In person or remote
Following one ICD/CRT-D shock Within 24 to 48 hours In person or remote
Following >1 ICD/CRT-D shocks Immediately In person
Any CIED until signs of battery depletion Annually In person
Any CIED following signs of battery depletion Every 1 to 3 months In person or remote
CIED: cardiac implantable electronic device; CRT: cardiac resynchronization therapy; ICD: implantable cardioverter-defibrillator; PPM: permanent pacemaker.
Adapted from: Wilkoff BL, Auricchio A, Brugada J, et al. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm 2008; 5:907.